Abstract
Expressions describing the absorbed dose and the number of incident photons necessary for the detection of a contrasting detail in X-ray transmission CT imaging of a circular phantom are derived as functions of the linear attenuation coefficients of the materials comprising the object and the detail. A shell of a different material can be included to allow simulation of CT imaging of the skulls of small laboratory animals. The equations are used to estimate the optimum photon energy in X-ray transmission computed microtomography. The optimum energy depends on whether the number of incident photons or the absorbed dose at a point in the object is minimised. For a water object of 300 mm diameter the two optimisation criteria yield optimum photon energies differing by an order of magnitude.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.