Abstract

A synchrotron X-ray diffractometer incorporating a pulsed field magnet for high fields up to 40 T has been developed and a detailed description of this instrument is reported. The pulsed field magnet is composed of two coaxial coils with a gap of 3 mm at the mid-plane for passage of the X-rays. The pixel detector PILATUS 100K is used to store the diffracted X-rays. As a test of this instrument, X-ray diffraction by a powder sample of the antiferromagnet CoO is measured below the Néel temperature. A field-dependent lattice distortion of CoO due to magnetostriction is observed up to 38 T.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.