Abstract

We have developed a new type of x-ray detector based on a superconducting transition-edge thermometer operated near 100 mK. A superconducting quantum interference device is used to measure the current through the thermometer, and negative electrothermal feedback is used to improve the energy resolution and shorten the thermal time constant. We have used a detector mounted on a scanning electron microscope to measure the energy of titanium Kα (4.5 keV) fluorescence x rays with a resolution better than 14 eV full width at half-maximum. Using two other devices, we have measured an energy resolution for Joule heat pulses of 2.6 eV at 1 keV and 0.2 eV at 4 eV, the best reported for any calorimeter. An electrical noise equivalent power of 3×10−18 W/√Hz was also measured, suggesting the use of these detectors as infrared bolometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.