Abstract

The X-ray detection capability of a scintillation detector equipped with a Cs2ZnCl4 single crystal was evaluated. The scintillation decay kinetics can be expressed as the sum of two exponential decay components. The fast decay component had a decay time constant of 1.8 ns, and its relative intensity was 95%. The total light output was 630 photons/MeV, and a subnanosecond timing resolution of 0.66 ns was obtained. The detection efficiency of 67.4 keV X-rays was 80% for a detector equipped with a 2.2-mm-thick Cs2ZnCl4 crystal. Thus, excellent timing resolution and high detection efficiency were achieved simultaneously.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.