Abstract

The aim of this study is to evaluate whether X-ray dark-field (DF) radiography is useful for the diagnosis of gout in birds and reptiles and whether this preclinical model could be helpful to establish this non-invasive imaging method in human medicine. A total of 18 limbs originating from 11 birds (7 different species) and 7 reptiles (4 different species) with and without suspected joint gout were measured using a grating-based X-ray dark-field setup and conventional X-ray examination, respectively. Each image acquisition generated a dark-field and a conventional absorption x-ray image. The results of the individual scans were compared with the results of a pathological examination and arthrocentesis. In 5 of the birds and 4 of the reptiles examined, gout was detected by pathologic examination. In each group, uric acid crystals were found in the joints of 3 animals by means of arthrocentesis. The uric acid crystals were detectable in 2 bird and 2 reptile limbs in the dark-field image. The study demonstrated that the urate crystals evoke a clearly visible dark field signal, whereas this was not the case in the conventional radiographs. The results obtained show that uric acid crystal detection using less invasive imaging methods in an animal model with birds and reptiles may expand gout diagnostics not only in veterinary medicine but also in human medicine and possibly replace arthrocentesis if a DF signal is detectable. Preclinical scanners which use X-ray dark-field and phase-contrast radiography already exist for hands and mammography.