Abstract

This paper illustrates the design, development and testing of a miniature mechanism to be integrated in endoscopic capsules for precise steering capabilities (Magnetic Internal Mechanism, MIM). The mechanism consists of an electromagnetic motor connected to a couple of small permanent magnets and immersed in a static magnetic field produced by an external permanent magnet or a by an electromagnetic coil. The overall steering capsule, integrating the magnetic steering mechanism and the vision system is 15.6 mm in diameter, 48 mm in length, 14.4 g in weight and can be oriented with an accuracy of 0.01 degrees . As regards system scalability, the capsule size could be reduced down to 11 mm in diameter by optimizing some mechanical components. On the other hand, the magnets size cannot be reduced because the magnetic link between internal and external magnets at typical operation distances (about 15 mm) would be weak.