Abstract

A wireless biotelemetry system for the transfer of digital data through intact skin and tissue has been developed to provide a safe and noninvasive means of communication between implanted medical devices and the outside of the body. The system utilizes 2 miniature infrared transmitter/receiver modules. Data are transmitted through intact skin and subcutaneous tissue on an 890 nm infrared carrier signal. The system has been evaluated in human cadavers and during in vivo implantation of artificial hearts and ventricular assist devices for durations of up to 96 h. Acceptable data transfer (error rate < 10(-5)) through a typical tissue thickness of 5-25 mm has been demonstrated. The ability to monitor and control a device from a remote site using public communication systems such as telephone lines and asynchronous transfer mode (ATM) systems has also been demonstrated. Design optimization is currently ongoing in preparation for clinical utilization with artificial heart systems and other implantable devices (such as rotary blood pumps).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.