Abstract

This in vitro study evaluated agreement among 10 trained evaluators when assessing implant stability with the Wired/Classic and Wireless/Periotest "M." A difference of 1 Periotest value (PTV) between the wired (-7) and wireless (-8) instruments was observed for the pretest calibration ring. No significant differences were found between the instruments and for all evaluators for all tests (analysis of variance, P < .05). Each instrument can provide meaningful and reproducible recordings of stability measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.