Abstract

Every operation could have a fire emergency, especially in the case of a tracheostomy. When a flammable gas meets a source of heat, the danger of fire is remarkable. A tracheal tube filled with a high concentration of oxygen is also a great risk factor for fire. Intra-tracheal tube fire is a rare, yet critical emergency with catastrophic consequences. Thus, numerous precautions are taken during a tracheostomy like, use of a special tube to prevent laser damage, ballooning of the tube with normal saline instead of air, and dilution of FiO2 with helium or nitrogen. Since the first recorded cases on tube fires, most of the fires were initiated in the balloon and the tip. In the present case report, however, we came across a fire incidence, which originated from the wire.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.