Abstract
In the paper a method of absolute measurements of continuous wave laser power is described. The influence of the apex angle and the walls smoothness on the reflection coefficient of the cone absorbers has been investigated. The construction and electrical circuit of a calorimeter containing a wire cone absorber are also described. This is designed for radiation measurements in the range from 25 μW to 25 mW.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.