Abstract

Objective: To explore window settings technology in the digital radiography of pneumoconiosis, and to improve the quality of the digital chest radiographs. Methods: 25 female workers in a asbestos product processing enterprise were examined with high-kV and DR chest radiographies on same day. Consistency of pneumoconiosis diagnosis results: were assessed. Results In the 75 groups of radiograph quality results, there are 62 groups that need to be processed window in order to make the correct diagnosis. The crude agreement among two chest radiographs was 28% for the profusion of small opacities and 76% for the classification of pneumoconiosis, the weighted kappa value (κ) was 0.33 (95%CI: 0.12-0.54) and 0.67 (95%CI: 0.42-0.92) . Conclusions: window settings technology can display different parts of chest and make DR chest radiograph to meeting the requirements of chest radiograph quality for pneumoconiosis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.