The calculation procedure of the parameters of a multielement transducer array for the optoacoustic tomography of biological objects with high spatial resolution values is proposed. A multielement transducer with given spatial resolution values in three dimensions has been developed based on the proposed procedure for the early detection of breast cancer. The transducer array consists of a set of 8 linear PVDF piezoelectric films located on a plane and a focusing cylindrical acoustic lens. A map of the transducer’s focal area and point spread function have been measured using the constructed transducer array. Spatial resolutions of the transducer array obtained experimentally are in agreement with their calculated values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.