Abstract

Longitudinal and radial tissue-fluid pressure distributions were determined beneath and adjacent to wide (12 and 18 cm) pneumatic tourniquet cuffs placed on intact human cadaveric arms and legs, respectively. Tissue fluid pressures exhibited relatively broad maxima at midcuff, and in most cases showed no differences at the various depths studied. Limb circumference was not a determining factor in the transmission of pressure to deeper tissue. We also investigated the effect of four cuff sizes (4.5, 8, 12, and 18 cm) on eliminating blood flow to the lower legs of normal subjects. The cuff pressure required to eliminate blood flow decreased as cuff width increased; thigh circumference was a determining factor in the pressure required to eliminate blood flow while using the smaller cuffs, but not while using the 18-cm cuff. Thus, wide cuffs transmit a greater percentage of the applied tourniquet pressure to deeper tissues than conventional cuffs; accordingly, lower cuff pressures are required, which may minimize soft-tissue damage during extremity surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.