ABSTRACT The mercury vapor monitor, MDC-1-01, described in this paper is the first prototype in the MDC series. It was designed and developed in our laboratory to measure airborne mercury vapor on a “continuous” basis in the range of 400 ng to 20 mg mercury per cubic meter of air. In comparison with other instruments, commercially available, it uses much improved principles as regards electronics, optics, and chemistry [l]. The double beam assembly, in combination with a selective filter, permits the measurement of mercury vapor in the presence of the most commonly encountered organic vapors and dusts. Practical applications are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.