Abstract

Robot-assisted total hip arthroplasty (RA-THA) helps with precise orientation of the prosthesis, but some RA-THA procedures are aborted intraoperatively and are converted to manual total hip arthroplasty (THA). This study aimed to analyse why RA-THA is sometimes aborted intraoperatively and to make recommendations accordingly. A total of 429 consecutive Mako THA cases in our prospective database from August 2018 to June 2021 were included in our study. All robotic procedures aborted intraoperatively for any reason were recorded. The patients' demographics, diagnoses, and surgeons' information were included in the statistical analysis to pinpoint the risk factors for intraoperative robot to manual conversion. Intraoperative RA-THA abortions occurred in 17 cases (3.96%) and the patients had to be converted to manual THA. The adverse events leading to intraoperative abortions included pelvic array loosening or malposition (5, 1.17%), inaccurate bone mapping or construction (6, 1.40%), inaccurate initial registration (4, 0.93%), and other reasons (2, 0.47%). Robot-related adverse events could be found in all perioperative steps of RA-THA, and some of these events might result in intraoperative abortion. Complex hip disease was a statistically significant factor for an increased risk of intraoperative abortion of RA-THA. Standardized surgical procedures and preoperative assessments can be helpful in reducing the rate of RA-THA abortions.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call