Abstract

Demonstration of a 3D version of the DREAM sequence (3DREAM) for rapid 3D flip angle and mapping of the human brain. A rectangular non-selective STE preparation is followed by a 3D readout with a Cartesian center-out spiral phase encoding order. This enables parallel imaging acceleration in both phase encoding dimensions as well as early capture of the prepared magnetization. mapping of the whole human head is demonstrated on a 7T system at a nominal resolution of 5 mm with and without parallel imaging acceleration. Artifacts caused by the different signal decay of the FID and STE signal during the long imaging train is suppressed by appropriate filtering of the FID image. Remaining blurring can be controlled by adjusting the echo train length and readout flip angle. 3DREAM provides a whole-brain flip angle map in a few seconds or individual maps for an 8-channel array in about a minute.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.