Whole-body vibration has been identified as a stressor to supine patients during medical transportation. The transmissibility between the input platform acceleration and the output acceleration of the head, sternum, pelvis, head-sternum, and pelvis-sternum of eight supine subjects were investigated. Vibration files were utilized in the fore-aft, lateral, and vertical directions. The power spectral density across the bandwidth of 0.5–20 Hz was approximately flat for each file. A comparison between a baseline rigid-support and a support with a long spinal board strapped to a litter has shown that the latter has considerable effects on the transmitted motion in all directions with a double magnification in the vertical direction around 5 Hz. The results also showed that the neck-collar has increased the relative head-sternum flexion–extension because of the input fore-aft vibration, but reduced the head-sternum extension–compression due to the input vertical vibration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.