Abstract

The ROTEM platelet device, a point-of-care whole blood platelet impedance aggregometer, is an add-on to the rotational thromboelastometry ROTEM delta device. The latter has been validated in dogs. We examined whether canine whole blood is suited for analysis with the ROTEM platelet device using adenosine-5'-diphosphate (ADP) and arachidonic acid (ARA) as agonists for platelet activation, and if there are significant differences between sample storage times and anticoagulants used. Subsequently, we determined canine reference intervals (RIs) for the ROTEM platelet device for ADP and ARA. In a pilot study, we examined whole blood from 7 dogs after 15-min and 60-min storage of lithium-heparinized samples and 40-min and 80-min storage of hirudinized samples. Statistical analysis showed no significant differences between ROTEM platelet device results for both ADP and ARA in lithium-heparin and hirudin anticoagulated canine whole blood. Lithium-heparinized blood samples analyzed after 15-min storage had the lowest coefficient of variation. RIs were determined for heparinized whole blood samples from 49 dogs after 15 min of storage.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.