We report hypercoagulability despite activated partial thromboplastin time (APTT)-guided heparin treatment during Berlin Heart-supported circulation in a 38-year-old man with heart failure for 19 days. The patient was anticoagulated using unfractionated heparin, acetylsalicylic acid and dipyridamole. Contact and tissue factor-activated thromboelastometry revealed increased clot firmness, although anticoagulation assessed by APTT was in accordance with the treatment protocol. Strength of polymerized fibrin was also increased. We saw no clinical signs of thrombosis. Thromboelastometry normalized after heart transplantation. Our results suggest that hypercoagulability is due to excess fibrin formation. Monitoring anticoagulation using APTT may, therefore, be misleading during mechanical cardiac assist.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.