Abstract

The results of comparative investigations into the optically stimulated luminescence (OSL) of anion-defective corundum excited by radiation of blue and white light-emitting diodes (LEDs) are reported. The continuous-wave OSL (CW-OSL) measurements showed that the white LED (compared to the blue LED) and the chosen geometry allow the amplitude of the OSL response to increase by a factor of more than 40 and the dosimetric information readout time to decrease by a factor of 20.The effect of deep traps on the behavior of the CW-OSL curves of anion-defective corundum was studied. It was found that the filling of deep traps essentially increases the TL yield with a dosimetric peak at 450 K and the CW-OSL-yield under stimulation with blue and white LEDs. It was also found that this filling causes a considerable change in the shape of the OSL curve and the time parameters of the decay.