Abstract

This study verified the resistance to displacement of six miniplate fixation methods after sagittal split osteotomy (SSO). SSO was performed in 30 polyurethane synthetic mandible replicas. The distal segments were advanced (4mm) and specimens were grouped according to the fixation method: four-hole standard miniplate; four-hole locking miniplate; six-hole standard miniplate; six-hole locking miniplate; six-hole standard sagittal miniplate; six-hole locking sagittal miniplate. Biomechanical evaluation was performed by applying compression loads to three points on the second molar region, using an Instron universal testing machine until a 3mm displacement of the segments occurred. Compression loads able to produce 3mm displacement were recorded in kN and subjected to analysis of variance (P<0.01) and Tukey's tests for comparison between groups (P<0.05). The locking sagittal miniplate showed higher resistance to displacement than the regular four- and six-hole locking and standard miniplates. No significant differences were observed between the locking sagittal miniplate and the regular sagittal or the four-hole locking miniplates. Two of the three groups with the best results had locking plate fixation methods. Fixation of SSO with a single miniplate is better accomplished using six-hole locking sagittal miniplates, six-hole standard sagittal miniplates, or four-hole locking miniplates; these methods are more resistant to displacement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.