Abstract
Here we explore the feasibility of using classical Mössbauer spectroscopy setup in scattering geometry to assess the elemental composition of samples using characteristic X-ray fluorescence. Even though, the scintillation detector of Mössbauer spectrometer yields rather low spectral resolution spectra, this can be effectively compensated using chemometric data processing. This hypothesis was first checked with simulated spectra of steel samples and afterwards was confirmed with real experiments. The results of Ti, Cr, Mn, Ni, Nb and Mo quantification in real steel samples obtained from Mössbauer spectrometer were compared with those obtained with commercial energy dispersive X-ray fluorescence (EDXRF) spectrometer. In case of Ti, Cr and Ni the accuracy attained with Mössbauer spectrometer was similar to that of an ordinary commercial EDXRF device.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.