Abstract
IntroductionThe cemented polyethylene cup has remained the standard acetabular implant for 50 years although there has been little research into cementing techniques. In the past, cement was previously inserted by sequential manual pressurisation (thumbing) but this technique was prone to contamination of the cement leading to weakening of fixation. In recent times, third-generation techniques using sealed pressurisation with rim preparation have been espoused with similar results. We were interested in establishing whether repeated cycles of compression of cement allowing adequate time for relaxation increases its depth of penetration, and the optimum period of relaxation required to achieve this goal.MethodA single mix of polymethylmethacrylate cement at dough stage was inserted into a model of the reamed acetabulum. Cyclical pressurisation of the cement with 50 N followed nine different patterns to simulate thumbing, constant pressure, and the application of a sealed and unsealed acetabular cup implant.ResultsA constant load was as effective as all variations of repeated cycles of load and relaxation except for 50 N pressure applied for four seconds with four second intervals. A four second interval of relaxation achieved significantly more penetration than five or three seconds.Following two minutes of constant pressure, the application of a sealed or unsealed thrust of the plunger had no effect on cement penetration.ConclusionThis study suggests that optimal polymethylmethacrylate cement penetration into the acetabulum occurs with cycled application of load for four seconds followed by four seconds of relaxation. The subsequent pressurisation with either a flanged or unflanged acetabular implant does not appear to improve cement penetration.
Highlights
The cemented polyethylene cup has remained the standard acetabular implant for 50 years there has been little research into cementing techniques
Following two minutes of constant pressure, the application of a sealed or unsealed thrust of the plunger had no effect on cement penetration
This study suggests that optimal polymethylmethacrylate cement penetration into the acetabulum occurs with cycled application of load for four seconds followed by four seconds of relaxation
Summary
The cemented polyethylene cup has remained the standard acetabular implant for 50 years there has been little research into cementing techniques. Cement was previously inserted by sequential manual pressurisation (thumbing) but this technique was prone to contamination of the cement leading to weakening of fixation. Third-generation techniques using sealed pressurisation with rim preparation have been espoused with similar results. We were interested in establishing whether repeated cycles of compression of cement allowing adequate time for relaxation increases its depth of penetration, and the optimum period of relaxation required to achieve this goal
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.