Abstract
Hemolysis is a cause of variability in test results for plasma potassium, LDH and AST and is a non-negligible part of measurement uncertainty. However, allowable levels of hemolysis provided by reagent suppliers take neither analytical variability (trueness and precision) nor the measurand into account. Using a calibration range of hemolysis, we measured the plasma concentrations of potassium, LDH and AST, and hemolysis indices with a Cobas C501 analyzer (Roche Diagnostics(®), Meylan, France). Based on the allowable total error (according to Ricós et al.) and the expanded measurement uncertainty equation we calculated the maximum allowable bias for two concentrations of each measurand. Finally, we determined the allowable hemolysis indices for all three measurands. We observed a linear relationship between the observed increases of concentration and hemolysis indices. The LDH measurement was the most sensitive to hemolysis, followed by AST and potassium measurements. The determination of the allowable hemolysis index depends on the targeted measurand, its concentration and the chosen level of requirement of allowable total error.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.