Abstract

BackgroundThe study summarised in this report aimed to investigate the interplay between fluence complexity, dose calculation algorithms, dose calculation spatial resolution and delivery characteristics (monitor units, effective field width and dose delivery against dose prediction agreement) was investigated. A sample set of complex planning cases was selected and tested using a commercial treatment planning system capable of inverse optimisation and equipped with tools to tune fluence smoothness.MethodsA set of increasingly smoothed fluence patterns was correlated to a generalised expression of the Modulation Index (MI) concept, in nature independent from the specific planning system used that could therefore be recommended as a predictor to score fluence "quality" at a very early stage of the IMRT QA process. Fluence complexity was also correlated to delivery accuracy and characteristics in terms of number of MU, dynamic window width and agreement between calculation and measurement (expressed as percentage of field area with a γ > 1 (%FA)) when comparing calculated vs. delivered modulated dose maps. Different resolutions of the calculation grid and different photon dose algorithms (pencil beam and anisotropic analytical algorithm) were used for the investigations.Results and Conclusioni) MI can be used as a reliable parameter to test different approaches/algorithms to smooth fluences implemented in a TPS, and to identify the preferable default values for the smoothing parameters if appropriate tools are implemented; ii) a MI threshold set at MI < 19 could ensure that the planned beams are safely and accurately delivered within stringent quality criteria; iii) a reduction in fluence complexity is strictly correlated to a corresponding reduction in MUs, as well as to a decrease of the average sliding window width (for dynamic IMRT delivery); iv) a smoother fluence results in a reduction of dose in the healthy tissue with a potentially relevant clinical benefit; v) increasing the smoothing parameter s, MI decreases with %FA: fluence complexity has a significant impact on the accuracy of delivery and the agreement between calculation and measurements improves with the advanced algorithms.

Highlights

  • The study summarised in this report aimed to investigate the interplay between fluence complexity, dose calculation algorithms, dose calculation spatial resolution and delivery characteristics was investigated

  • Of notice that the Zxy component is dominating over the other two. This is due to the fact that in the optimisation process, no smoothing is applied in this cross direction but the importance of x-y fluence discontinuities cannot be ignored when delivery accuracy issues are to be considered and 2D evaluations like the γ pass/fail analysis are performed

  • Fixing Modulation Index (MI) = 19, the probability to have %FA < 5% would be: 90%, 40%, 60% and 22% respectively. These results suggest that, with this value of MI, only for the case of Analytical Algorithm (AAA) and 2.5 mm it would be possible to guarantee a satisfactory agreement between calculation and measurement

Read more

Summary

Introduction

The study summarised in this report aimed to investigate the interplay between fluence complexity, dose calculation algorithms, dose calculation spatial resolution and delivery characteristics (monitor units, effective field width and dose delivery against dose prediction agreement) was investigated. The optimisation process is a computational problem, potentially susceptible to noise and artefacts (high frequency spatial fluctuations) producing sharp fluence peaks and valleys in millimetric spatial scale. These features could translate into difficult patterns for the delivery system, prolonged beam-on time and increased sensitivity to all conventional treatment uncertainties. The authors demonstrated that, as a trade-off, the more 'complex' the intensity patterns, the higher the number of monitor units (MU) will be to deliver the prescribed doses This could affect, due to linac potential limitations, both the quality and accuracy of delivered doses

Objectives
Methods
Results
Conclusion