Currently, there is significant interest in quantifying the performance of digital radiography systems. Digital radiography systems can be thought of as continuous linear shift-invariant systems followed by sampling. This view, along with the large number of pixels used for flat-panel systems, has motivated much of the work which attempts to extend figures of merit developed for analog systems, in particular, the detective quantum efficiency (DQE) and the noise equivalent quanta (NEQ). A more general approach looks at the system as a continuous-to-discrete mapping and evaluates the signal-to-noise ratio (SNR) completely from the discrete data. In this paper, we study the effect of presampling blur on these figures of merit. We find that even in this idealized model the DQE/NEQ formulations do not accurately track the behavior of the fully digital SNR. Therefore, DQE/NEQ cannot be viewed as indicators of the signal-known-exactly/background-known-exactly (SKE/BKE) detectability. In order to make design decisions by optimizing detectability one must work with the fully digital definition of detectability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.