Abstract
Bead shedding has been reported in both hip and knee arthroplasties and can result in increased polyethylene wear and metal ion release. The rates of bead loss were recorded for each device type, and the interfacial bead-substrate shear load to failure and contact areas of the beads were determined. The 3 components with the highest rate of bead loss, namely, S-Rom, Omnifit, and Optifix acetabular shells, recorded the lowest shear loads to failure. Percentage bead-substrate contact area ranged between 7% and 37% and was the most important factor affecting bead shedding. Other factors influencing bead loss in decreasing order of importance were bead fusion, bead size, component type (acetabular or knee), component loosening, and time in situ.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.