Abstract

In emergency cases, rapid extracorporeal membrane oxygenation (ECMO) device initialization is able to drastically reduce the incidence of patient morbidity and/or mortality. Pre-assembled and ready-to-use ECMO circuits might save up to 30-60 critical minutes in patient management. Six ECMO circuits (Oxygenator D905 EOS with REVOLUTION™ pump and Sorin PTS) were assembled in the operating room in standard conditions and then placed at 37°C for 35 days in order to evaluate possible contamination and ingrowth of micro-organisms. Every 7 days after ECMO circuit assembly and wet-priming, samples of priming fluid were analyzed to verify the presence/absence of possible common contaminants (Enterobacteriaceae, Staphylococcus aureus and fungi). Moreover, two supplementary circuits, used as positive controls, were deliberately inoculated with a known concentration of a Escherichia coli strain and prime samplings carried out at different time-points to determine bacterial growth rate. Sterility was maintained in the ECMO circuits for up to 35 days.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.