Abstract

The unloaded quality factor of the cavity resonator is the ratio between the stored energy of the cavity resonator to the power loses in the cavity resonator. The homemade rectangular cavity resonator in X-band shows higher unloaded quality factor compare with standard cavity resonator in the TE102 mode. Because the inner walls of rectangular cavity resonator are treated through high quality polishing and high purity Au plating. Also the inner walls are made by printed circuit board which has thin Cu foil, two problems such as mechanical vibration and thermal expansion can be solved by minimizing unwanted eddy current. Through the ferromagnetic resonance measurement by using our rectangular cavity resonator, we can be obtained reasonable values of resonance frequency and linewidth by using NiFe thin film. As a result, the Gilbert damping constant from the experimental result is in good agreement with the typical value of damping parameter of the NiFe thin film.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.