Abstract
A well-type detector installed in the Modane underground Laboratory (LSM) can combine both low background and high detection efficiency and it is well suited for the analysis of small amounts of environmental samples. Reference materials such as IAEA-447 (moss-soil), IAEA-RG-Th1 and IAEA-RG-U1 were used for the detector calibration, owing to a chemical composition close to those of the environmental samples. Nevertheless, the matrix effects and the true coincidence summing effects must be corrected from the full energy peak efficiency (FEPE). The FEPE was performed for a wide range of energy by a semi-empirical method using Monte Carlo simulation (MCNP6), intended for environmental measurements such as lake sediments dating. In the well geometry, the true coincidence summing effects could be very important and correction factors have been computed in three different ways.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.