Abstract

This study was designed to describe the accuracy and intrasession reliability of weightbearing measurements obtained with digital bathroom scales during comfortable bilateral upright stance in patients with stroke. The 20 subjects were all independently ambulatory with a unilateral assistive device. Their weightbearing during comfortable stance was measured over three trials. Analysis of variance revealed that the sum of the weightbearing measurements of the paretic and nonparetic sides did not differ significantly from total body weight. Thus, the measurements were considered accurate. Some variability in the weightbearing measurements was demonstrated with analysis of variance, which revealed significant differences between trials on both sides. Nevertheless, reliability as described by intraclass correlation coefficients was good (0.829 paretic side, 0.876 nonparetic side). Although such reliability is sufficient to justify clinical use, methods for increasing reliability should be examined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.