Abstract

The dependence of the wedge factors (WFs) on field size (FS) and depth for a fast neutron beam has been investigated. In a previous study (Popescu et al 1999 Med. Phys. 26 541), a method was presented that allows a simple and accurate way of calculating the wedge-factor dependence on FS and depth in the case of a photon beam. The validity of a similar approach is tested in the present study for neutron beam dosimetry. The clinical neutron therapy system at the University of Washington (UW) has a flattening filter assembly consisting of two filters: a small field filter and a large field filter. Despite this complication, the approach presented in Popescu et al (1999 Med. Phys. 26 541) can be used to describe the WF dependence on FS and depth (d).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.