Abstract
Background: The thermal effect correlated with implant osteotomy could produce significant effects on the healing process and fixture osseointegration. The aim of the present investigation was to assess the heat generation and surface wearing of dental implant drills and manual tappers during simulated osteotomies on animal ribs. Methods: Steel drills (20 units per type) and tappers (20 units per type) were evaluated for a total of 30 osteotomies. The infrared thermal analysis was performed at the first and thirtieth osteotomy. The surface alteration and wearing was assessed by energy dispersive spectroscopy–scanning electron microscopy (EDS-SEM) prior to and after use. Conclusions: The drill material produced a non-significant temperature change during bone osteotomy. Lower heating was reported for manual tappers in favor of a manual osteotomy instead rotary instruments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.