Abstract
Four patients in whom the S-ROM Total Hip System Polydial polyethylene liner was used illustrate the importance of, and difficulties in, detecting polyethylene wear-through before a complete acetabular revision is necessary. The patients in these cases had full thickness polyethylene wear-through 11 or more years after total hip arthroplasties. Dorr and Livermore wear measurements from radiographs of the patients underestimated the actual wear and did not indicate wear-through. Because excessive head penetration was not evident radiographically with the S-ROM components, the cases showed the necessity of using additional measures to diagnose wear-through, particularly with these components. Repeat radiographs at a higher kilovoltage were used to better visualize the outline of the femoral head, and hip aspiration was used to diagnose metallosis. These cases also showed the importance of yearly followup evaluations, especially 10 or more years after surgery for patients with the S-ROM Polydial locking mechanism.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.