Abstract

We present a patient with a metal and polyethylene wrist prosthesis who had recurrent synovial swelling on the volar side of the wrist. Surgical exploration showed synovitis, a severely worn and broken polyethylene insert, polyethylene wear particles, and titanium debris from fretting corrosion. Pathological analysis showed macrophage-phagocytosing polyethylene wear particles and titanium oxide debris. This shows that polyethylene and titanium wear in wrist prosthesis might be a bigger problem than originally believed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.