Abstract

Forty-seven explanted Porous Coated Anatomic (PCA, Howmedica, Rutherford, NJ) cementless acetabular components were acquired at revision surgery. All the components articulated against CoCrMo femoral heads of 32-mm diameter. The penetration depth and angle were measured using the shadowgraph technique. The wear volume was then calculated using Kabo's formula. Using weighted linear regression analysis, the mean penetration rate and mean volumetric wear rate were calculated to be 0.23 (SE, 0.03) mm 3/y and 96 (SE, 13) mm 3/y, respectively. The creep component was not found to be significantly different from zero. The clinical wear factor, k clinical, for this cohort was also calculated using linear regression analysis but with the assumption that creep was zero. The value found, k clinical = 1.93 (SE, 0.29) × 10 −6 mm 3/N-m, was similar to those in previous studies involving cemented joints with a 22-mm femoral head diameter. The similar k clinical values of these substantially different joint types suggest that the high volumetric wear rate for the PCA joint can be attributed entirely to its larger head size and the younger, more active, patient profile. Fixation technique and metal backing seem not to influence the rate of wear.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.