Abstract

Most of today’s artificial joints rely on an articulating couple consisting of a CoCrMo alloy and a medical grade polyethylene. The wear of the polyethylene component is the major cause for long-term failure of these prostheses since the wear debris leads to adverse biological reactions. The polyethylene wear is usually measured by gravimetric methods, which are limited due to a low sensitivity and accuracy. To demonstrate the reliability of ultra thin layer activation (UTLA) as an alternative technique, wear tests on a cross-linked ultra-high-molecular weight polyethylene (XLPE) sliding against CoCrMo were performed on a wear tester featuring multi-directional sliding motion. The amount of polyethylene wear was evaluated by both UTLA and gravimetry. The particular TLA method used in this work employed the implantation of 7Be radioactive recoils into the polyethylene surface by means of a light mass particle beam. The results indicate that apart from its relatively high sensitivity, UTLA also offers the possibility for on-line measurements of polyethylene wear. This makes it a viable and complementary technique in wear test studies for medical implant purposes especially for those involving wear resistant materials and for rapid wear screening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.