Abstract

AbstractWear behaviour of the artificial joint replacements plays a critical role in implant failure. The UHMWPE‐AISI 316L tribological system, widely used on joint prostheses, was studied. Two surface modifications were applied to the stainless steel counterface: DC‐pulsed plasma nitriding and duplex (DC‐pulsed plasma nitriding + TiN coating by cathodic‐arc evaporation) treatments. An Amsler Disc Machine Type A‐135 was used as an approximation to real loading conditions. Worn surfaces, weight loss curves and wear debris were analysed. The results show that the nitriding and duplex treatments improve the wear resistance of the tribological system. Different types of damages were observed on the UHMWPE worn surfaces depending on the surface treatment of the stainless steel.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.