Abstract

Purpose: To investigate the use of a system using EM tracking, postprocessing and error-detection algorithms for measuring brachytherapy catheter locations and for detecting errors and resolving uncertainties in treatment-planning catheter digitization. Methods: An EM tracker was used to localize 13 catheters in a clinical surface applicator (A) and 15 catheters inserted into a phantom (B). Two pairs of catheters in (B) crossed paths at a distance 4.5 mm (A) and >5.5 mm (B); errors were detected for shifts on average >2.0 mm (A) and >2.4 mm (B). Both mix errors associated with undistinguishable catheter artifacts were detected and at least one of the involved catheters was identified. Conclusion: We demonstrated the use of an EM tracking system for localization of brachytherapy catheters, detection of digitization errors and resolution of undistinguishable catheter artifacts. Automatic digitization may be possible with a registration between the imaging and the EM frame of reference. Research funded by the Kaye Family Award 2012.« less