Abstract
Purpose: To provide case studies for computed radiographic(CR) system image exposure measurement metrics and drive manufacture innovation. Methods and Materials:Computed radiographyimage plate exposure indices were monitored over time. For entries in the exposure log file that indicated more than five times of the usual value, time stamps in the entries were matched manually to images stored on PACS to identify possible causes. Results: Large variation in the values of the exposure index was found. Both operator errors and software failures were found for images of high value of exposure index. After corrective actions, the FWHM of the distribution of the exposure index has decreased although the range of the value remains the same. Cases from Pediatric Intensive Care Unit are the most difficult to handle because of the patient size and medical complications. Conflict of Interest (only if applicable): JL is an employee of Agfa HealthCare Corporation
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.