Abstract
We adopt an orthogonal wavelet packet decomposition (OWPD) filtering approach to cancel harmonic interference noises arising from an AC power source in time domain and remove the resulting rf voltage interference noise from the mass spectra acquired by using a charge detection frequency-scan quadrupole ion trap mass spectrometer. With the use of a phase lock resampling technique, the transform coefficients of the rf interference in signals become a constant, exhibiting a shift of the baseline in different rf phases. The rf interference is therefore removable by shifting the baselines back to zero in OWPD coefficients. The approach successfully reduces the time-domain background noise from 1367 electrons (rms) to 408 electrons (rms) (an improvement of 70 %) and removes the high frequency noise components in the charge detection ion trap mass spectrometry. Unlike other smoothing or averaging methods commonly used in the mass-to-charge (m/Ze) domain, our approach does not cause any distortion of original signals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society for Mass Spectrometry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.