Abstract

The effective pixel size of a two-dimensional wavelength-shifting-fibre (WLS-fibre)-based neutron image detector was improved from 0.5 to 0.16 mm by including a fibre-optic taper (FOT) between the scintillator screen and the WLS fibre. The WLS-fibre-based detector consisted of a thin ZnS/ 6LiF screen, a FOT and WLS ribbons crossed in the x and y directions. The demonstrator detector had 16 fibre channels in each direction, and the light signals in each fibre were read out individually. The FOT was constructed from fine glass fibres with a taper ratio of 3.1 and served as an image magnifier. The prototype detector equipped with the FOT exhibited a spatial resolution of 0.3 mm, compared to the spatial resolution of 0.8 mm of the original detector without the FOT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.