Abstract

Phase interrogation surface plasmon resonance (P-SPR) biosensors have the highest sensitivity among different types of surface plasmon resonance (SPR) biosensors. However, P-SPR sensors have small dynamic detection range and complex device configuration. To solve these two problems, we designed a multi-channel P-SPR imaging (mcP-SPRi) sensing platform based on a common-path ellipsometry scheme. A wavelength sequential selection (WSS) technique for P-SPRi sensing is developed to select the optimal sensing wavelengths according to different refractive indexes (RIs) of the samples, so the inconsistency of SPR signal response for different biomolecule types caused by the small dynamic detection range is eliminated. And a dynamic detection range of 3.7×10-3 RIU is achieved, which is the largest among the current mcP-SPRi biosensors. Remarkably, the individual SPR phase image acquisition time has been greatly reduced to 1s by using WSS method instead of whole spectrum scanning, which enables the high-throughput mcP-SPRi sensing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.