Abstract

A pulsed distributed feedback quantum cascade laser (QCL) operating near 957 cm-1 was employed in wavelength modulation mode for spectroscopic trace gas sensing applications. The laser was excited with short current pulses (5-10 ns) with < 2% duty cycle. The pulse amplitude was modulated with a linear sub-threshold current ramp at 20 Hz resulting in a ~ 2.5 cm-1 frequency scan, which is typically wider than what has been reported for these lasers, and would allow one to detect molecular absorption features with line widths up to 1 cm-1. A demodulation approach followed by numerical filtering was utilized to improve the signal-to-noise ratio. We then superimposed a sine wave current modulation at 10 kHz onto the 20 Hz current ramp. The resulting high frequency temperature modulation of the distributed feedback (DFB) structure results in wavelength modulation (WM). The set-up was tested by recording relatively weak absorption lines of carbon dioxide. We demonstrated a minimum detectable absorbance of 10-5 for this spectrometer. Basic instrument performance and optimization of the experimental parameters for sensitivity improvement are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.