Abstract

AbstractAn x‐ray fluorescence spectrometric method is described for the determination of elements present in the fumes evolved during welding. Synthetic standards were prepared using salts of Si, Mn, Cr, Fe, Pb, Cu, Al and Ni for evaluation. The standards were mixed with boric acid in the ratio 1 : 1 and briquettes were prepared by the double pellet method. The samples were collected in filter‐paper discs by exposing them during welding. The average standard deviation of the method is about 7%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.