Abstract
A database of 416 wavelength- and energy-dispersive EPMA measurements on tilted specimens of NiAl, TiO(2) and Ti(3)Al is presented. The analyses were performed between 10 and 30 kV and the tilt angles were varied between 0 degrees and 60 degrees in seven steps. The necessary hardware modifications for the specimen holder are discussed, as well as the various focusing techniques used in the measurements. A comparison between the experimental data, the calculations of our proza96t program and the results of Monte Carlo simulations shows that up to 50 degrees tilt the predictions of our software are more than satisfactory. At larger tilt angles some deviations become noticeable. The Monte Carlo simulations appear to produce deviations at a somewhat earlier stage already, for reasons as yet unknown.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.