Abstract

The effect of imperfect collimation on the fringe separation of a Fizeau wavemeter is calculated. It is shown that illumination in a plane normal to the wedge face and parallel to the wedge axis enables the detector array to be positioned such that the wavelength error caused by spherical wave-front curvature introduced by the beam expander becomes insignificant to 1 part in 10(8), and no dispersion correction for the wedge plate is required. It is also shown that a sealed air wedge is capable of a wavelength precision approaching 1 part in 10(8).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.