Abstract

This study demonstrates the adverse effects of water in exhaled breath samples on the accuracy of breath biomarker analysis when using gas chromatography. The presence of water in exhaled breath significantly modifies the retention times and peak areas of compounds, particularly for low-boiling, early eluting compounds. To tackle this issue, a two-step approach is introduced. The process begins with thorough desorption of the sorbent tube using a high split ratio and a short analysis duration, followed by a secondary analysis of the same tube. The efficacy of the new, straightforward approach was illustrated using humid breath samples and 57 compound standard mixture. This study highlights the importance of proper sample pretreatment and analysis to ensure reliable and accurate results in clinical research.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.