Abstract

It has recently been shown that the use of strong organic solvents as diluent is possible in RPLC, provided that the solvent used as diluent is retained more strongly by the column than the analytes in the sample. In this study, the phenomenon was further studied experimentally using several water-immiscible solvents (ethyl acetate, isopropyl acetate, and methyl isobutyl ketone) and several model analyte compounds. In all cases, analyte peak distortion was minimal provided the analyte eluted earlier than the diluting solvent, in contrast to substantial broadening and distortion when the analyte eluted after the diluting solvent. The potential analytical utility of this approach is discussed, and an example of a practical application is also presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.