The dosimetry of carbon-ion beams based on calibrated ionization chambers (ICs) still shows a significantly higher uncertainty compared to high-energy photon beams, a fact influenced mainly by the uncertainty of the correction factor for the beam quality k Q . Due to a lack of experimental data, k Q factors in carbon-ion beams used today are based on theoretical calculations whose standard uncertainty is three times higher than that of photon beams. To reduce their uncertainty, in this work, k Q factors for two ICs were determined experimentally by means of water calorimetry for the spread-out Bragg peak of a carbon-ion beam, these factors are presented here for the first time. To this end, the absorbed dose to water in the 12C-SOBP is measured using the water calorimeter developed at Physikalisch-Technische Bundesanstalt, allowing a direct calibration of the ICs used (PTW 30013 and IBA FC65G) and thereby an experimental determination of the chamber-specific k Q factors. Based on a detailed characterization of the irradiation field, correction factors for several effects that influence calorimetric and ionometric measurements were determined. Their contribution to an overall uncertainty budget of the final k Q factors was determined, leading to a standard uncertainty for k Q of 0.69%, which means a reduction by a factor of three compared to the theoretically calculated values. The experimentally determined values were expressed in accordance with TRS-398 and DIN 6801-1 and compared to the values given there. A maximum deviation of 2.3% was found between the experiment and the literature.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call